5
Settings
............................
5.4.3.1 – 5.4.3.2
5.4.3.1 Accept data
Procedure:
Accept the reference values by pressing the <Measurement key>.
The new data are stored.
The display shows Return in reverse video.
Press the <Measurement key> several times to return to the main menu.
5.4.3.2 Reject data
Procedure:
Select the menu line Reject data by turning the <Control ball> and confirm the action by pressing
the <Measuring key>.
The new data will not be stored; stored data are retained.
The display shows the menu line Return in reverse video.
Press the <Measurement key> several times to return to the main menu.