X-Rite D19C Standard > D19C Full Upgrade Manuale di Servizio Pagina 49

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 164
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 48
Densitometric functions
............................
4
4.12
4.12 Hue error
........................................................................................................................................
Select paper white or absolute white base under Settings k Measurement parameters k
White base.
........................................................................................................................................
Procedure:
Select the menu line Hue error by turning the <Control ball>.
Activate the selected function by pressing the <Measurement key>.
Select the menu line Measure paper white by turning the <Control ball>.
Position the measuring aperture on the paper white and initiate a measurement by pressing the
<Measurement key>.
The display White measurement done is shown for about 1 second.
Position the measuring aperture on the color patch to be measured and initiate a measurement by
pressing the <Measurement key>.
The hue error H is shown on the display.
Select the menu line Return by turning the <Control ball> and press the <Measurement key> to
return to the main menu.
Vedere la pagina 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 163 164

Commenti su questo manuale

Nessun commento