Message Message on Error source
No. display
23 Not calibrated Missing calibration values
24 Temperature failure The instrument’s
temperature exceeded
the limit of 50°C during
the charging of the
battery. The ambient
temperature was to high.
(25) Measure different color The same color was
measured two times in
the trapping function.
(26) Measure solid tones first In the Auto functions, an
overprint field has been
measured before the solid
fields.
(27) Battery charging The battery will be
approx. 3h charged in less then
3 hours
(28) Battery charging The battery will be
approx. 15h charged in approx.
15 hours with low power
GretagMacbeth chargers
(12V DC, 450mA)
29 No battery The rechargeable battery
is not connected
(32) Battery empty The battery is discharged;
no further operations are
possible
(33) Battery full The battery is fully
charged
7.7
Reméde
Calibrate instrument
Charge the battery at
room temperature
conditions
Perform the correct
measuring sequence:
Measure 1st printed color
Measure 2nd printed color
Measure overprint
Measure first the
appropriated solid fields.
Connect rechargeable
battery
Connect charger
Disconnect charger
Vedere la pagina 131