Settings
............................
5.4.3
5.4.3 Entry or correction of reference and tolerance values
........................................................................................................................................
After entering reference or tolerance values, the current values are indicated in the respective
entry fields.
........................................................................................................................................
Procedure:
Select the menu line References by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.
The display shows Reference input in reverse video.
Activate the function by pressing the <Measurement key>.
The display shows Measure paper white in reverse video.
Select the menu line of the desired color by turning the <Control ball> and press the <Measurement
key>.
The entry field for the entering or correction of the reference value appears in reverse video on the
display.
Enter the new reference value or correct the current value and terminate the entry by pressing the
<Measurement key>.
The entry field for the entering or correction of the tolerance value is shown in reverse video on the
display.
Enter the new tolerance value or correct the current value and terminate the entry by pressing the
<Measurement key>.
The display shows the menu line with the newly entered values in reverse video.